Evaluation of magnetic field homogeneity using in-out signal cycle mapping in gradient recalled echo images of a mixed water/oil phantom as a rough indication for daily quality control
Abstract
Objective: Magnetic field (B0) homogeneity is important for the performance of a magnetic resonance imaging (MRI) system. Traditionally, B0 homogeneity was measured using the spectral peak or phase-mapping methods. However, these procedures are not generally accessible to the MRI operator and are rarely performed routinely. This study proposes a novel method for measuring B0 homogeneity that can be implemented in daily quality control (QC).
Methods: When a uniformly mixed water/oil phantom was imaged using a gradient recalled echo (GRE) pulse sequence, the signal intensity dynamically changed with echo time (TE). From this, the resonant frequency was calculated with a simplex curve-fitting algorithm on a pixel-by-pixel basis. The standard deviation of resonant frequency (SD) was used as the index of B0 homogeneity. The appropriate TE pattern and feasibility of B0 homogeneity evaluation were examined.
Results: Over seven TEs (choosing nominal in-phase, out-phase, and the midpoints of both) were required to measure stable SD in a 1.5-T scanner. As B0 homogeneity worsened, the SD became larger at the off-center position. Although a positive correlation was observed with the width of the spectral peak obtained by the phase-difference method, the SD value was about 5 × 104 times greater. Therefore, SD can be used only as an index of B0 homogeneity. Similar results were obtained using a 0.3-T scanner. A map and SD can be obtained by acquiring several GRE images of a water/oil mixed phantom within a few minutes.
Conclusions: In-out signal cycle mapping can be easily implemented for daily QC in all MRI scanners.
Full Text:
PDFDOI: https://doi.org/10.5430/jbgc.v6n2p7
Journal of Biomedical Graphics and Computing ISSN 1925-4008 (Print) ISSN 1925-4016 (Online)
Copyright © Sciedu Press
To make sure that you can receive messages from us, please add the 'Sciedupress.com' domain to your e-mail 'safe list'. If you do not receive e-mail in your 'inbox', check your 'bulk mail' or 'junk mail' folders